Loading...

    Explore
    • Policy Monitoring
    • Live Policy Feed
    • Billing Codes
    • Healthcare Providers
    • Health Insurance Payers
    • States
    • Service Lines
    • Taxonomy Codes
    Lookup tools
    • Billing Code Lookup
    • CPT Modifier Lookup
    • Provider Taxonomy Code Lookup
    • Healthcare Provider & Facility Lookup
    • Health Insurance Payer Lookup
    • Payer Contact Directory

    Policy & Benchmark Tools

    • Prior Authorization Lookup
    • Coverage Criteria Lookup
    • Timely Filing Lookup
    • Billing Code Reimbursement Rates
    • Payer Comparison Dashboard
    • Facility Comparison Dashboard
    • All tools

    Company

    • About OpenPayer
    • Data Sources
    • Request a Policy / Feedback
    • Trek Health
    OpenPayerby Trek Health
    LinkedInYouTube

    © 2026 Trek Health. All rights reserved.

    OpenPayer is built and maintained by Trek Health.

    v1.0.32
    OpenPayerPolicy UpdatesClinical/Medical PolicyBluecross Idaho Neuropathic Pain Coverage | OpenPa…
    Ask Trek AI about this page
    CurrentBluecross IdahoPolicy MP 3.03.07

    Audio-Visual Neuromodulation for Neuropathic Pain

    Customize your policy alerts

    Sign up for Bluecross Idaho Policy MP 3.03.07 alerts

    Get alerted when Policy MP 3.03.07 changes without checking for updates manually.

    Monitor payer policy activity

    Receive email alerts as soon as relevant payer policies are published or updated.

    This policy governs coverage determinations for FDA-authorized prescription audiovisual digital therapeutics used to manage neuropathic pain for providers and members subject to applicable Blue Cross of Idaho benefit plans.

    Policy Summary
    PayerBluecross Idaho
    PolicyAudio-Visual Neuromodulation for Neuropathic Pain
    Policy CodePolicy N/A
    Change TypeNo material change
    Effective DateSep 1, 2026
    Last ReviewMay 28, 2026
    Next Review DateN/A
    Key ActionTreat FDA-authorized prescription audiovisual digital therapeutics for neuropathic pain as investigational.
    SourceLink
    POLICY UPDATE CHANGES

    No material clinical or coverage changes in this revision.

    InvestigationalCoverage status
    3Scope criteria
    6–12 monthsControlled follow-up
    Single trialEvidence base
    1

    OpenPayer is powered by Trek Health's payer performance platform. Trek continuously ingests, validates, and normalizes Transparency in Coverage data alongside payer policies and other commercial payer data to create a structured payer intelligence foundation. OpenPayer uses this foundation to deliver personalized search results, dynamically generated policy pages, and tailored policy monitoring based on each user's payers, specialties, billing codes, and areas of interest. The same intelligence powers broader payer performance workflows, including reimbursement benchmarking, contract evaluation, payer negotiations, and financial decision-making.

    FDA-authorized therapeutic

    Coverage Criteria

    Scope of Reviewed Technologies

    This review is restricted to prescription audiovisual therapeutics for management of neuropathic pain that meet ALL of the following scope criteria. The reviewed products deliver their primary therapeutic effect through integrated hardware-software systems using visual and/or auditory stimulation and may have received FDA De Novo classification or 510(k) clearance, depending on predicate-device availability.

    ALL of the following

    • The technology is a prescription digital therapeutic (PDT): software that performs one or more medical purposes as the primary therapeutic component of a hardware-software system, excluding software used solely to store or transmit medical information.
    • The technology has received marketing clearance or approval from the U.S. Food and Drug Administration (FDA).
    • The technology is prescribed by a healthcare provider.

    Investigational Evidence Standard

    A determination that an investigational technology improves net health outcomes requires ALL of the following minimum evidence elements:

    ALL minimum evidence elements

    • Independent replication of the pivotal findings, with prognostic variables controlled as covariates.
    • A prospectively established minimum clinically important difference (MCID) derived from an external cohort independent of the pivotal trial.
    • Controlled follow-up of at least 6 to 12 months to establish durability of effect.>= 6 to 12 months

    Evidence Sufficiency

    The evidence is insufficient to determine that the technology improves the net health outcome. The current evidence does not satisfy ALL of the following sufficiency considerations:

    Current evidence sufficiency considerations

    • Evidence includes independent replication beyond the single pivotal double-blind, sham-controlled randomized controlled trial (N=75).

      The available evidence is limited to one trial with no independent replication. In that trial, the prespecified primary endpoint—between-group neuropathic pain symptom severity at week 14, after treatment—was met (mean difference 8.75; 95% CI 1.30 to 16.20; p=.021). Both groups improved at week 10 with no significant between-group difference; only the active group maintained improvement at week 14 while the sham group trended toward baseline. Active participants also had significant reductions in use of multiple pain-medication classes compared with sham.

    • The clinically important treatment effect is established using a prospectively established MCID validated in an external cohort independent of the pivotal trial.

      The responder analysis used an anchor-derived clinically important difference threshold drawn from the same trial data rather than an externally validated MCID. The active response rate was 67% versus 32% for sham (NNT 2.89; chi-square=4.44; p=.035).

    • The analysis controls for baseline prognostic variables as covariates.

    Coding and Outcome Measures

    FDA Product Codesmixed
    QYNFDA Product Code
    FDA Clearancemixed
    DEN250005De Novo
    CPTCPT
    ..
    HCPCSHCPCS
    E1399Durable medical equipment, miscellaneous
    ICD-10-CMICD-10
    G89.21Chronic pain due to trauma
    G89.22Chronic post-thoracotomy pain
    G89.28Other chronic postprocedural pain
    G89.29Other chronic pain
    G89.3Neoplasm related pain (acute) (chronic)
    G89.4Chronic pain syndrome
    M79.2Neuralgia and neuritis, unspecified
    Numeric Rating Scale for Pain Intensity
    MeasureNumeric Rating Scale for Pain Intensity (0–10)
    Clinically meaningful improvementAt least a 2-point difference or at least a 30% reduction from baseline
    Assessment formatSingle-item self-report of current pain intensity, from 0 (no pain) to 10 (worst imaginable pain)
    Neuropathic Pain Symptom Inventory
    MeasureNeuropathic Pain Symptom Inventory (NPSI), 0–100 total score
    Established MCID

    Provider Actions and Coverage Impact

    Denial Risk

    Investigational technology

    Use of U.S. Food and Drug Administration-authorized prescription audiovisual digital therapeutics for neuropathic pain is considered investigational.

    Denial Risk

    Sana Device evidence is insufficient

    Coverage may be denied because the evidence is insufficient to determine that the Sana Device improves net health outcomes. The evidence consists of one pivotal trial without independent replication, an internally derived and externally unvalidated MCID, and important baseline prognostic imbalances not controlled in the primary analysis.

    Denial Risk

    Investigational evidence standard not met

    The technology remains investigational unless future evidence provides all of the following: independent replication with prognostic variables controlled as covariates; a prospectively established MCID derived from an external cohort independent of the pivotal trial; and controlled follow-up of at least 6 to 12 months to establish durability of effect.

    Clinical Background

    Neuropathic pain results from a lesion or disease of the somatosensory nervous system and is estimated to affect 7% to 10% of the general population. Common symptoms include burning, pressing, tingling, and electrical sensations, which may occur spontaneously or with hyperalgesia or allodynia. The condition can substantially impair physical function, psychological well-being, and quality of life.

    Available management includes pharmacologic therapies such as gabapentinoids, tricyclic antidepressants, serotonin-norepinephrine reuptake inhibitors, topical agents, and opioid-like medications. Behavioral and psychological therapies may be used as adjuncts. Because medication effectiveness is often moderate and symptoms may persist, audiovisual digital therapeutics are intended to provide an adjunctive or alternative nonpharmacologic treatment option.

    Definitions

    Prescription Digital Therapeutic (PDT)
    Primary therapeutic componentSoftware that performs one or more medical purposes as the primary therapeutic component of a hardware-software system
    ScopeIncludes prescription digital therapeutics using integrated hardware-software systems for therapeutic purposes
    ExclusionDoes not include software used solely to store or transmit medical information
    Sana Device
    Device typePrescription wearable audiovisual stimulation mask
    StimulationDelivers personalized pulses of light and sound to the eyes and ears
    Intended therapeutic use
    Policy Summary
    PayerBluecross Idaho
    PolicyAudio-Visual Neuromodulation for Neuropathic Pain
    Policy CodePolicy N/A
    Change TypeNo material change
    Effective DateSep 1, 2026
    Last ReviewMay 28, 2026
    Next Review DateN/A
    Key ActionTreat FDA-authorized prescription audiovisual digital therapeutics for neuropathic pain as investigational.
    SourceLink
    On This Page
    Coverage CriteriaCoding and Outcome MeasuresProvider Actions and Coverage ImpactClinical BackgroundDefinitions

    Baseline imbalance across age, sex, pain severity, and pain duration was not controlled for in the primary model.

  1. Controlled follow-up establishes durability of effect for at least 6 to 12 months.>= 6 to 12 months

    It is unknown whether neuroplastic changes observed at week 14 are sustained, whether repeated treatment cycles are needed to maintain benefit, or what retreatment intervals, if any, would be appropriate.

  2. Evidence demonstrates consistent improvement across relevant secondary outcomes, including mood, anxiety, sleep quality, and quality of life.

    No significant between-group differences were observed for these secondary outcomes.

  3. No universally established anchor-based MCID has been published for the NPSI total score
    Preliminary estimatesDistributional estimates range from approximately 6 to 10 points across four methods; anchor-based and etiology-specific validation is still needed
    Brief Pain Inventory–Short Form Pain Interference
    MeasureBrief Pain Inventory–Short Form pain interference composite score
    Clinically meaningful improvementAt least a 1-point improvement on the pain interference composite score
    Instrument scopeThe questionnaire assesses pain severity and interference with daily functioning
    Patient Global Impression of Change
    MeasurePatient Global Impression of Change, a 1–7 scale
    Responder thresholdA rating of 6 (much improved) or 7 (very much improved)
    Assessment formatSingle-item self-report of perceived overall change since treatment initiation
    SF-36
    Measure36-Item Short Form Health Survey (SF-36), scored 0–100
    Clinically meaningful improvementAt least 5 to 10 points on the Physical Component Summary or Mental Component Summary
    DomainsGenerates Physical and Mental Component Summary scores for health-related quality of life
    Hospital Anxiety and Depression Scale
    MeasureHospital Anxiety and Depression Scale, with anxiety and depression subscales scored 0–21 each
    Clinically meaningful improvementAt least a 1.5-point improvement per subscale
    SubscalesIncludes HADS-A for anxiety and HADS-D for depression; scores of at least 8 indicate probable caseness
    Pittsburgh Sleep Quality Index
    MeasurePittsburgh Sleep Quality Index, scored 0–21
    Clinically meaningful improvementAt least a 3-point improvement
    Assessment scopeA 19-item questionnaire measuring sleep quality over the prior month across seven components
    Pain Visual Analog Scale
    MeasurePain Visual Analog Scale (Pain-VAS), using a 0–10 cm continuous scale
    Clinically meaningful improvementAt least a 1.5 to 2.0 cm reduction
    Assessment formatSingle-item continuous assessment of pain intensity
    Patient Health Questionnaire-9
    MeasurePatient Health Questionnaire-9, scored 0–27
    Clinically meaningful improvementAt least a 5-point improvement
    Assessment scopeNine-item self-report of depressive symptom frequency over the prior 2 weeks; scores of at least 20 indicate severe depression
    Generalized Anxiety Disorder 7-item Scale
    MeasureGeneralized Anxiety Disorder 7-item Scale, scored 0–21
    Clinically meaningful improvementAt least a 4-point improvement
    Assessment scopeSeven-item self-report of anxiety symptom frequency over the prior 2 weeks; scores of at least 15 indicate severe anxiety
    Controlled Follow-up
    Controlled follow-upAt least 6 to 12 months after treatment completion
    PurposeEstablish durability of the treatment effect
    Outcomes monitoredSymptoms, functional outcomes, quality of life, and treatment-related morbidity
    • Independent replication with prognostic variables controlled as covariates
    • Prospectively established MCID derived from an external cohort independent of the pivotal trial
    • Controlled follow-up of at least 6 to 12 months
    Denial Risk

    Coverage impact

    The available evidence is insufficient to determine that the technology improves net health outcomes; therefore, the technology remains investigational.

    Provides sensory neuromodulation for neuropathic pain and is intended for home or clinic use under a healthcare provider’s direction
    Sana Device: Description and Therapeutic Mechanism
    Session protocolUses a 16-minute treatment session protocol
    Therapeutic mechanismPersonalized light and sound stimuli modulate neural activity through sensory neuromodulation targeting central mechanisms of neuropathic pain
    PersonalizationProprietary software personalizes stimulus delivery using heart-rate-variability biofeedback
    Neuropathic Pain Symptom Inventory (NPSI)
    InstrumentValidated self-reported questionnaire assessing neuropathic pain severity
    Symptom dimensionsAssesses burning, pressing, electric shock, tingling, and pain evoked by brushing, pressure, and cold
    StructureThe NPSI is a 12-item questionnaire assessing qualitative and quantitative neuropathic pain across five symptom dimensions
    MCID and NPSI Context
    MCID meaningMinimum clinically important difference
    NPSI statusNo universally established anchor-based MCID has been published for the NPSI total score
    Preliminary NPSI contextA January 2026 systematic meta-analysis reported distributional estimates of approximately 6 to 10 points; etiology-specific and anchor-based validation is still needed
    Outcomes of Interest
    Follow-up timingFollow-up after treatment and at 6 to 12 months after treatment completion is relevant
    Clinical outcomesMonitor symptoms, functional outcomes, and quality of life
    Safety and burdenMonitor treatment-related morbidity
    MCID
    DefinitionA minimum clinically important difference threshold derived from an anchor
    UseUsed to identify clinically meaningful improvement
    Evidence limitationFor the Sana Device evidence base, the threshold and the result claiming success were derived from the same pivotal trial data
    FDA De Novo Authorization
    Authorization pathwayFDA De Novo authorization
    DeviceSana Device
    Authorization dateJanuary 2026